Advantages

  • High accuracy, dual angle measurement (90° + 25°), wide measurement span and optional with colour measurement
  • Values and measurement curves indicated directly on point of measurement
  • Proven measurement technology, measurement method identical to laboratory instrument LabScat 2
Download brochure

Technical data

Wavelength turbidity
LED 650 nm
Measuring principle turbidity
90°/25° scattered light measurement
Nominal range turbidity
0 … 1000 EBC (0 … 4000 NTU)
Resolution turbidity
0.001 EBC of full scale
Reproducibility
±1 % @ 0 .. 2 EBC
±2 % @ 2 .. 100 EBC (90°)
±3 % @ 2 .. 100 EBC (25°)
±10 % @ 100 .. 1'000 EBC
Repeatability
0.001 EBC bzw. ±0.25%
Linearity
±0.5% full scale @ 0 .. 2 EBC
Reaction time
less than 2 s (step response - limit value switch)
Wavelength color (optional)
LED 430 nm
Nominal range color
0 .. 50 EBC (fix path)
Enclosure
Stainless steel 1.4301
Protection class
IP 66
Operation voltage
24 VDC ± 10 %
Warm-up time
3 min.
Dimensions
Ø 100.5 x 128 mm
Weight
Appr. 2.3 kg
Sample temperature
-10 .. +100 °C
120 °C max. 2 h
150 °C max. 1 h
Sample pressure
- Max. 4 MPa (40 bar) → Sensor head with sapphire windows
- Measurement cell → Please note the specification, standard VARINLINE®-housing 1 MPa (10 bar)
Connections
DN 40 … DN 150, 1 1/2“ ... 6“ VARINLINE® -housings
Ambient environment
-10 .. +50 °C
+55° C -> max. temperature environment +85° C
+60° C -> max. temperature environment +75° C
0 .. 100 % relative humidity
Cleaning
CIP/SIP compatible up to 120 °C
Material sensor head

Stainless steel 1.4404

Windows material
Sapphire
Download data

Downloads

Preview Title Type

Download