AquaMaster
Advantages
-
Compact, modular complete system
-
Unrestricted parameter combination
-
Central, integrated control unit with colour touch screen
-
Strong, reliable brand names: SIGRIST & HAMILTON
-
One contact for all measurements
-
Possibility to integrate external sensors (e.g. ColorPlus)
-
Easy calibration and maintenance, no tools required
Technical data
- Interfaces
- Ethernet, Modbus TCP, micro SD-Card
- Display
- ¼ VGA, 3.5“
- Operation voltage
- 100 .. 240 VAC, 47 .. 63 Hz or
18 .. 30 VDC - Power consumption
- max. 10 W
- Dimensions
- approx. 55 x 115 x 40 cm (B x H x T)
- Protection class
IP66 (Electronics)
- Outputs
- 4 x 0/4 .. 20 mA, galvanic separated
2 x Relays 250 VAC, 4 A
5 x digital outputs, free configurable - Sample temperature
- 0 .. +40 °C
- Ambient temperature
- 0 .. +50 °C
- Ambient humidity
- 0 .. 100% rel. F.
- Sample pressure
- max. 0.6 MPa (6 bar)
- Operation
- Touchscreen
- Inputs
- 1 x for optional flow meter
2 x 0/4 .. 20 mA
4 x digital outputs, free configurable - Options
- Current output module, 4 x 0/4 .. 20 mA
Profibus DP
Modbus RTU - Turbidity measurement
- AquaScat, model selectable (see information for the corresponding instrument model)
- Sensor type pH
- Measuring principle: Potential measurement against reference
- Measuring units: pH, Temperature [°C, K, °F]
- Measuring range: pH 0 .. 14
- Material in contact with medium: Glass, FPM (Viton), Electrolyte (Polisolve Plus), Reference (Everref-L)
- Sensor type Conductivity
- Measuring principle: 4-Pole Measurement
- Measuring units: µS/cm, mS/cm, Temperature [°C, K, °F]
- Measuring range: 1 .. 300'000 µS/cm
- Accuracy: ± 3% @1 µS/cm .. 100 mS/cm, ± 5% @100 .. 300 mS/cm
- Material in contact with medium: Stainless steel 1.4435/316L, Ra 0.4 µm (N5), PEEK (FDA approved), EPDM (FDA approved)
- Sensor type Dissolved Oxygen
- Measuring principle: Optical (Luminescence)
- Measuring units: µg/l, mg/l, ppm, ppb, %-sat, %-vol, Temperature [°C]
- Measuring range:4 ppb .. 25 ppm
- Accuracy (@25°C): 1±0.05 %-vol, 21±0.2 %-vol, 50±0.5 %-vol
- Material in contact with medium: Stainless steel 1.4435/316L, Silicone (FDA approved), EPDM (FDA approved)
- Sensor type Redox/ORP
- Measuring principle: Potential measurement against reference
- Measuring units: mV, Temperature [°C, K, °F]
- Measuring range: -1500 .. 1500mV
- Material in contact with medium: Glass, FPM (Viton), Platinum